4339B
Functional
T
est
11.
Press the
IN5
switch
on the
Handler Interface
T
ester
,
and
conrm
that
the
4339B
is
triggered.
Handler
Interface Output
Test.
12.
Start the
handler interface
output test
using
the
following
procedure:
a.
Press
to
display the
conguration setting
menu.
b.
Select
more
using
or and
press
.
c.
Select
Svc
using
or and
press
.
d.
Select
HNDL
using
or and
press
.
13. Set
the
LED
PW
switch
on the
Handler Interface
T
ester
to
ON
.
Caution
Do
not
set
the
LED
PW
switch
to
ON
,
except
when
doing
the
handler
interface
output
test.
The
LEDs
may
cause
the
4339B
to
shut-down
during
normal
operation.
14.
Conrm that
the LEDs
on
the
Handler
Interface
T
ester
light
in
the
order
shown
in
Figure
9-12,
in accordance
with
the
4339B
display
.
Figure 9-12. Handler Interface
Output Order
15. Set the
LED PW
switch on the Handler Interface Tester to OFF.
16. Press
twice to exit from the test mode.
Maintenance 9-23