4339B
T
able A
-4. O
VLD/Over-Current/N.C.
Operations(Meas
.
parameter:
I)
Display
Handler
Output
GPIB
Output
1
Solutions
Data
Mode
Comprtr
Mode
<
stat
>
2
:1
O
VLD
O
VLD
HIGH /HI
<
data
>
:
9.9E37
Select
an
appropriate
measurement
range.
(Overload)
<
comp
>
3
:
2
/HI
<
stat
>
2
:4 Reduce the
source
voltage
.
Over-Current O
V
CURR
HIGH &
<
data
>
:
9.9E37
Check
whether
a short
circuit is
occurred.
/NOT
READY
<
comp
>
3
:
2
<
stat
>
2
:
2
Cancel
the
bad
contact
between
DUT and
N.C. N.C. N.C. /NO
CONT
A
CT
<
data
>
:
9.9E37
contact-pin.
(No-Contact)
<
comp
>
3
:8
O
VLD
<
stat
>
2
:
3
Select
an appropriate
measurement range
.
& N.C. N.C. /HI
<
data
>
:
9.9E37
Cancel
the
bad
contact
between
DUT
and
N.C.
<
comp
>
3
:
2
contact-pin.
O
VLD
/HI
<
stat
>
2
:
5
Select
an
appropriate
measurement
range
.
& O
V
CURR
HIGH &
<
data
>
:
9.9E37
Reduce
the
source
voltage
.
Over-Current /NOT
READY
<
comp
>
3
:
2
Check
whether
a
short
circuit
is
occurred.
1
Refer
to
:FETCh?
in
Chapter
5
2
Measurement
status (0:Normal,
1:Overload, 2:No-Contact,
4:Over-Current)
3
Comparison
result
(1:In,
2:High,
4:Low
,
8:No-Contact)
A-4 Manual Changes