EasyManuals Logo

Agilent Technologies 6470 User Manual

Agilent Technologies 6470
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
6 Reference
Positive Ion Mode Checkout Conditions
108 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
Sample (stock) Reserpine, 5
ng/µL
Reserpine, 5
ng/µL
Reserpine, 5
ng/µL
Reserpine, 5
ng/µL
Reserpine, 5
ng/µL
Concentration after
dilution
500 fg/µL 500 fg/µL 500 fg/µL 500 fg/µL 500 fg/µL
Dilution solvent 70:30
acetonitrile:
water without
formic acid
70:30
acetonitrile:
water without
formic acid
70:30
acetonitrile:
water without
formic acid
70:30
acetonitrile:
water without
formic acid
70:30
acetonitrile:
water without
formic acid
Total sample amount
injected on column
1 pg 1 pg 1 pg 1 pg 1 pg
Sample order number
(p/n)
G2423A G2423A G2423A G2423A G2423A
Tabl e 1 4 MS parameters (continued)
ESI (6410/6420/
6430/6460)
AJS (6460) AJS (6470) AJS (6490) AJS (6495)