Installation Completion and System Verification 3
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 69
9 Run the Positive Ion Mode checkout worklist. Repeat step 2 starting on
page 63, with these changes:
• Run the appropriate Positive Ion Mode checkout worklist.
• Use reserpine instead of chloramphenicol.
• Load each solvent blank and reserpine sample data file into the
Qualitative Analysis program.
• Generate Extracted Ion Chromatograms of the ion product (195.1 m/z for
Positive Ion Mode).
Figure 36 Example of a Positive mode signal-to-noise report.