Installation Completion and System Verification 3
Step 5. Verify the Triple Quadrupole LC/MS sensitivity
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 69
9 Run the Positive Ion Mode checkout worklist. Repeat step 2 starting on
page 63, with these changes:
Run the appropriate Positive Ion Mode checkout worklist.
Use reserpine instead of chloramphenicol.
Load each solvent blank and reserpine sample data file into the
Qualitative Analysis program.
Generate Extracted Ion Chromatograms of the ion product (195.1 m/z for
Positive Ion Mode).
Figure 36 Example of a Positive mode signal-to-noise report.