76 Chapter2
Making Adjustments
10. CAL YTF Adjustment Routine
14.Press the following keys to store the mixer bias value for band 4 in
nonvolatile memory.
CAL, More 1 of 4, More 2 of 4
SERVICE CAL
EXECUTE TITLE
15.Press the following keys to check the default mixer bias DAC values.
PRESET
CAL, More 1 of 4, More 2 of 4
SERVICE DIAG
DISPLAY CAL DATA
Confirm that the “Optimum Bias” DAC values displayed for band 1
through band 4 are 1600, 20, 1300, and 520.
16.Repeat the CAL YTF adjustment procedure.
17.Perform the CAL MXR adjustment procedure.