3-27
Replacing Assemblies
A100 Display, Backlights, and Filter
Display Removal
CAUTION: The new flat panel display comes with a protective plastic sheet over
the glass. Remove this plastic very slowly to avoid damage due to ESD.
CAUTION: The surface of the display is very easily scratched. Avoid touching it
with your bare hands or other objects. Use a blower to remove any dust from the
display surface.
1 Remove the rubber mount that houses the display assembly from the
front frame
2 Pull open the flaps of the rubber mount.
3 Remove the display shield board (1).
4 Now it is possible to carefully lift up on the display (2) to remove it
from the rubber mount.