126 Agilent Cary 8454 UV-Visible Spectroscopy System Installation Guide
7 LAN Communication, Installation, Connection and Configuration
Using the Agilent BootP Service
5 To edit the BootP Settings go to Start >All Programs > Agilent BootP
Service > Edit BootP Settings. The BootP Settings screen appears.
When this screen opens the first time, the default settings from
installation are displayed.
6 To edit the TabFile, select Maintain bootp tabfile?
The default tab file was created at installation and is located at
Program Files\Common Files\Agilent Shared\BootP\bin\TabFile. It
contains configuration information entered on this screen.
If necessary you can
change the BootPtab File Location by using the browse button on the
right (a valid BootP tabfile must exist).
create your own template for the tabfile by selecting Create template
BootP tabfile? and click Create Template.
7 Select Do you want to log bootp requests?
The default log file was created at installation and is located at
Program Files\Common Files\Agilent Shared\BootP\bin\LogFile. It
contains an entry for every time a device requests configuration
information from BootP.
If necessary you can change the BootPlog File Location by using the
browse button on the right (a valid BootP tabfile must exist).
8 Perform your necessary changes by e.g. creating a log file entry of a
new instrument and edit the new instrument using the Launch Manager.
9 Deselect Do you want to log bootp requests?
10 Click OK to save the values or Cancel to discard them. The program
ends.
11 To restart the BootP Service go to Start > Settings > Control Panel >
System and Security and select Administrative Tools > Services. The
Service screen appears.
12 Right click Agilent BootP Service.
13 Select Start.
14 Close the Services and Administrative Tools screen.
15 This completes the configuration.