Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
173
Front-Panel Key and SCPI Command
Reference
4.4.1.2 Meas Setup
Accesses the Meas Setup menu.
4.4.1.2.1 Tuning
Sets tuning mode and center frequency for manual tuning mode.
4.4.1.2.2 Gate Time
Sets the gate time for frequency counter.
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FCOunter:TFRequency <freq>
[:SENSe]:FCOunter:TFRequency?
[:SENSe]:FCOunter:TAUTo 0|1|OFF|ON
[:SENSe]:FCOunter:TAUTo?
Unit: Hz|kHz|MHz|GHz
Preset: 1 GHz
State Saved: Saved in instrument state.
Min: 100 kHz
Max: Hardware dependent on the PSA specified model number
Example: :FCO:TFR 100 MHz
:FCO:TFR?
:FCO:TAUT OFF
:FCO:TAUT?
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FCOunter:GLENgth <time>
[:SENSe]:FCOunter:GLENgth?
[:SENSe]:FCOunter:GAUTo 0|1|OFF|ON
[:SENSe]:FCOunter:GAUTo?
Unit: s | ms | us
Preset: 100.0ms
State Saved: Saved in instrument state.
Min: 1us
Max: 500.0ms