EasyManuals Logo

Agilent Technologies E4443A Manual

Agilent Technologies E4443A
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #216 background imageLoading...
Page #216 background image
Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
216
Front-Panel Key and SCPI Command
Reference
4.4.5.3 Trace / View
Allows you to select the desired measurement view from the following selections:
Numeric results
Waveform result
4.4.5.4 Meas Setup
Displays the measurement setup menu for the FM Deviation Measurement when the FM Deviation
Measurement key has been selected in the Measure menu.
4.4.5.4.1 IF BW
Used to specify the IF BW, if the IF BW Type is set to Man.
Example: :DISP:FMD:WIND:TRAC:Y:COUP OFF
:DISP:FMD:WIND:TRAC:Y:COUP?
Mode: MRECEIVE
Key Path:
View/Display
Preset: NUMeric
State Saved: Saved in instrument state.
Range: Numeric results | Waveform result
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FMDeviation:BANDwidth|BWIDth <freq>
[:SENSe]:FMDeviation:BANDwidth|BWIDth?
Preset: 1 MHz
State Saved: Saved in instrument state.
Range: 1 kHz to 10 MHz
Notes: When IF BW Type is set to “AUTO”, the measurement does not use this
value.
Example: :FMD:BAND 1MHz
:FMD:BAND?

Table of Contents

Other manuals for Agilent Technologies E4443A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4443A and is the answer not in the manual?

Agilent Technologies E4443A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4443A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals