Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
216
Front-Panel Key and SCPI Command
Reference
4.4.5.3 Trace / View
Allows you to select the desired measurement view from the following selections:
• Numeric results
• Waveform result
4.4.5.4 Meas Setup
Displays the measurement setup menu for the FM Deviation Measurement when the FM Deviation
Measurement key has been selected in the Measure menu.
4.4.5.4.1 IF BW
Used to specify the IF BW, if the IF BW Type is set to Man.
Example: :DISP:FMD:WIND:TRAC:Y:COUP OFF
:DISP:FMD:WIND:TRAC:Y:COUP?
Mode: MRECEIVE
Key Path:
View/Display
Preset: NUMeric
State Saved: Saved in instrument state.
Range: Numeric results | Waveform result
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:FMDeviation:BANDwidth|BWIDth <freq>
[:SENSe]:FMDeviation:BANDwidth|BWIDth?
Preset: 1 MHz
State Saved: Saved in instrument state.
Range: 1 kHz to 10 MHz
Notes: When IF BW Type is set to “AUTO”, the measurement does not use this
value.
Example: :FMD:BAND 1MHz
:FMD:BAND?