Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
240
Front-Panel Key and SCPI Command
Reference
4.4.7.2.4 Avg Number
Sets the number that will be averaged. After the specified number (average counts) has been averaged, the
averaging mode (termination control) setting determines the averaging action.
4.4.7.2.5 Fast Mode
Allows you to toggle the
Fast Mode between On and Off. Fast Mode can speed up the measurement.
Mode: MRECEIVE
Key Path:
Meas Setup
Remote Command: [:SENSe]:MODRate:AVERage:COUNt <integer>
[:SENSe]:MODRate:AVERage:COUNt?
[:SENSe]:MODRate:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:MODRate:AVERage[:STATe]?
Preset: 25, ON
State Saved: Saved in instrument state.
Min: 1
Max: 8192
Example: :MODR:AVER:COUN 10
:MODR:AVER:COUN?
:MODR:AVER:STAT OFF
:MODR:AVER:STAT?
Key Path:
Meas Setup
Remote Command: [:SENSe]:MODRate:FAST 0|1|OFF|ON
[:SENSe]:MODRate:FAST?
Default: OFF
State Saved: Saved in instrument state.
Range: ON|OFF
Example: :MODR:FAST ON
:MODR:FAST?