EasyManua.ls Logo

Agilent Technologies EasyEXPERT

Agilent Technologies EasyEXPERT
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
26 PowerDiode
26.2 If-Vf: Diode forward bias characteristics, SMU voltage pulse (A.05.00)
[Supported Analyzer]
B1505A
[Description]
Measures the forward bias voltage vs current characteristics. SMU voltage pulse is used for the forward bias output.
[Device Under Test]
Junction device, diode, 2 terminals
[Device Parameters]
Temp: Temperature
[Test Parameters]
Memo: Memorandum
Anode: SMU connected to Anode terminal, primary sweep voltage output
VfStart: Sweep start voltage for Anode terminal
VfStop: Sweep stop voltage for Anode terminal
VfLinearStep: Sweep step voltage for Anode terminal, effective if Scale=LINEAR
Cathode: GNDU connected to Cathode terminal
PulsePeriodMode: Pluse period mode, AUTO or MANUAL
ManualPulsePeriod: Pulse period, effective if PulsePeriodMode=MANUAL
PulseWidth: Pulse width
IfLimit: Current compliance
Scale: Scale of sweep, LINEAR, LOG10, LOG25, or LOG50
[Extended Test Parameters]
PulseBase: Pulse base voltage
PulseAvgCnt: Pulse averaging count
HoldTime: Hold time
IfZero: Y axis (If) minimum value
MeasurementTime: Actural measurement time for a pulse period
[Measurement Parameters]
Forward current If
[User Function]
Ta: Temperature Ta=Temp
[X-Y Plot]
For Scale=LINEAR:
X axis: Forward voltage Vf (LINEAR)
Y1 axis: Forward current If (LINEAR)
Y2 axis: Forward current If (LOG)
For Scale=LOG10, LOG25, or LOG50:
X axis: Forward voltage Vf (LOG)
Y1 axis: Forward current If (LINEAR)
Y2 axis: Forward current If (LOG)
[List Display]
Forward voltage Vf
Forward current If
[Parameter Display Area]
Temperature Ta=Temp
Agilent EasyEXPERT Application Library Reference, Edition 8
26-5