74 Agilent 1200 Series Autosampler Reference Manual
3 Troubleshooting and Test Functions
Needle Down Failed
The needle arm failed to move down into the needle seat.
The lower position of the needle arm is monitored by a position sensor on the
sampling unit flex board. The sensor detects the successful completion of the
needle movement to the needle seat position. If the needle fails to reach the
end point, or if the sensor fails to recognize the needle arm movement, the
error message is generated.
Probable Causes
Needle installed incorrectly, or wrong needle type (too long).
Defective or dirty position sensor.
Defective motor.
Sticking spindle assembly.
Defective ASM board.
Suggested Actions
Ensure the correct needle type is used, and installed correctly.
Exchange the sampling unit flex board.
Exchange the needle drive motor.
Exchange the spindle assembly or sampling unit assembly.
Exchange the ASM board.

Table of Contents

Other manuals for Agilent Technologies G1314B VWD

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1314B VWD and is the answer not in the manual?

Agilent Technologies G1314B VWD Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1314B VWD
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals