Verification and Calibration Appendix B
Series N6700 User’s Guide 143
Test Record Form – Agilent N6734A and N6744A
Agilent N6734A and N6744A Report No _______________ Date __________________
Description Model Minimum Specs. Results Maximum Specs.
Constant Voltage Tests
Voltage Programming & Readback
Minimum Voltage Vout
Front Panel Display Readback
High Voltage Vout
Front Panel Display Readback
Both
Both
Both
Both
− 15 mV
Vout − 35 mV
34.93 V
Vout − 70 mV
__________
__________
__________
__________
+ 55 mV
Vout + 35 mV
35.07 V
Vout + 70 mV
CV Load Effect
Both
− 4 mV
__________
+ 4 mV
CV Source Effect
Both
− 4 mV
__________
+ 4 mV
CV Ripple and Noise
peak-to-peak
rms
Both
Both
N/A
N/A
__________
__________
+ 15 mV
+ 5 mV
Transient Response
Voltage at 200 µs
Both
− 200 mV
__________
+ 200 mV
Constant Current Tests
Current Programming & Readback
Minimum Current Iout
Front Panel Display Readback
High Current Iout
Front Panel Display Readback
Both
Both
N6734A
N6744A
N6734A
N6744A
− 10 mA
Iout − 4 mA
1.47775 A
2.9755 A
Iout − 6.25 mA
Iout − 8.5 mA
__________
__________
__________
__________
__________
__________
+ 30 mA
Iout + 4 mA
1.52225 A
3.0245 A
Iout + 6.25 mA
Iout + 8.5 mA
CC Load Effect
Both
− 2 mA
__________
+ 2 mA
CC Source Effect
Both
− 1 mA
__________
+ 1 mA
Instrument Settings
Test Description N6734A N6744A
Voltage Programming & Readback, Zero Voltage 20 mV, 1.5 A 20 mV, 3 A
Voltage Programming & Readback, High Voltage 35 V, 1.5 A 35 V, 3 A
CV Load Effect, Source Effect, Ripple and Noise 35 V, 1.5 A 35 V, 3 A
Transient Response 35 V, from 0.75 A to 1.5 A 35 V, from 1.5 A to 3 A
Current Programming & Readback, Zero Current 10 mA, 35 V 10 mA, 35 V
Current Programming & Readback, High Current 1.5 A, 35 V 3 A, 35 V
CC Load Effect, Source Effect 1.5 A, 35 V 3 A, 35 V