Function Ab Ke Printout
AVERAG. MODE
AM
F2
01:NiCr +0123.4 -0012.0 +0000.0 °C 1.0000 E+0 STSTOP
CROSS SECTION
CS
F2
CROSS SECT: 01: 00078 cm2 DIAMETER: 00100 mm
DIAMETER
DN
F2
CROSS SECT: 01: 00078 cm2 DIAMETER: 00100 mm
ANALOG START
AS
F2
ANALOGSTART:01: +0000.0 °C
ANALOG END
AE
F2
ANALOGEND: 01: +0100.0 °C
ATM. PRESSURE
DEVICE ADDR.
(device
programming)
mb
A
F2
F2
DEVICE: G00 M11 A01 P03/11/00
ATM PRESS: +01013. mb
CJ TEMP: +0023.5 °C
U SENSOR: ! 12.5 V
HYSTERESIS: 10
CONFIG: FC-DA--- -L--
ALARM: -1-3
A1: DK0 Un
A2: AK1
s. man. 6.2.5
BAUD RATE
(sensor
programming)
BR
F2
AMR ALMEMO 2290-4 s. man. 6.2.3
MS RANGE LVMAX LVMIN BASE D FACTOR EXP AVG COMMENT
01:Ntc +040.00 - - - - - °C 1.0123 E+0 - - - Temperature
02:°o H +0060.0 +0020.0- - - %H - - - E+0 - - Humidity
MEASCYCLE: 00:00:00 S0251.9 F0104.7 E W010 C--U-
PRINTCYCLE: 00:10:00 Un 9600 bd
LOCKING
(extended sensor
programming)
LM
F2
MS ZEROPT SLOPE VM K FUNC EOFSET EFACT ANA-ST ANA-END B1 MX EF AH AL ZF RPM
01:+0000.0 +1.0000 5. 1 MESS +00000 32000 +0000.0 +1000.0-01 M1 -- S- E2 05 12.0
s. man. 6.10.1
9.4 Device Address and Networking
All ALMEMO
®
instruments can be very easily networked to centrally acquire the
measured values of several instruments that are located at different places (s.
man. 5.3). For communicating with networked devices it is mandatory that each
device has its own address as only one device is allowed to respond to each
command. Therefore, before any network operation it is necessary that all
connected devices are set to different device numbers. The function DEVICE
ADDRESS ´A´ is used for this purpose. It can be selected with the key F2 and,
at first, and the currently set device number is displayed, which is usually
factory-set to 00. It can then be modified by the normal data entry (s. 5.4).
Function DEVICE ADDRESS ´A´
Selection with key F2: ... Input see 5.4
F2
0 1 A
Example:
Address 01
Only successive numbers between 01 and 99 should be entered for network
operation so that the device 00 cannot be falsely addressed in case of a power
supply failure.
44 ALMEMO
2290-4
Digital Data Output