12. Function menus
47
can be entered directly (see 9.5). Once started the timer counts down to the
next cycle.
Function
Cycle timer
:
Cycle timer : 00:02:00 S
Cycle (hh:mm:ss), Saving ON, List format
Switch memory activation ON / OFF by pressing : <MON/MOFF>
The output format in which measured values are output via the interface can
be set by pressing <FORM> or in function
Output format
(see. 12.5.8.3). (Print
layouts see Manual 6.6.1)
Modify format by pressing <FORM>
Format, adjacent columns ´n´:
Cycle timer: 00:02:00Sn
To modify format <FORM>
Format, table ´t´:
Cycle timer: 00:02:00St
To start cyclic measuring point scan press <START>
The following symbols will be displayed as verification in the status bar contin-
uously, i.e. so long as the measuring operation is running. (see 9.2) :
The start arrow lights up (continual) ´ ´
Lights up when data is being output via the interface (continual) ´COM´
Appears when measured values are saved (continual) ´REC ´
To stop cyclic measuring point scan : <STOP> ´ll ´
12.5.6 Numbering of measuring operations
To identify measuring operations or series of measuring operations these can
be individually numbered before starting. This number is output or saved when
the next measuring point scan starts. In this way individual measuring opera-
tions can be assigned to certain types of measurement or certain measuring
points (see Manual 6.7).
After selecting the function
Number
the 6-digit number is entered as normal
(see 9.5). You can use digits 0 to 9 and also the characters A, F, N, P, and - or
_ (space). The number is activated as soon as it has been entered; it will then
be followed by the letter ´A´ until the next cyclic or manual measuring operation
is saved.
Number function : (e.g. room 12, measuring point 1) Number: 12-001 A
To zero-set and deactivate the number press <CLR>
To activate and deactivate the number press : <ON> , <OFF>
To increment and activate the number press < +1 >
12.5.7 Memory space, memory output, clearing the
memory
When measured values are being recorded the
Memory capacity free
function
continuously displays the memory capacity still available. Selecting this function
enables two softkeys, one for direct memory output and one for memory clear-
ing. The output format is as set in the cycle (see 12.5.5, 12.5.8.1, and Manual
6.6.1).