This guide does not replace the user manual
Intended for all the persons authorized to adjust settings
Quick guide
It is recommended to carry out this test after each
patient circuit and / or expiratory valve replacement.
Press Autotests
Close the patient circuit, validate to start the testing
sequence (<2 min).
After the beep and the message, remove the cap to
end the tests.
• Autotests
Contact
Air Liquide Medical Systems
Parc de Haute Technologie
6, rue Georges Besse
92 182 Antony Cedex - France
Phone : +33 (0)1 40 96 66 00
Hotline : 0820 14 63 59
• Expiratory valve MONNAL EVA
Extraction : press the eject
button
Single use
valve
Autoclavable
valve
Cleaning protocol (see user manual)
Assembly instructions
It is recommended to use a bacteriological filter on
the inspiratory output, to limit the risk of retro-
contamination (see user manual).
• Bacteriological filter
Assembly :
• Integrate the silicone disks
into the membrane (1)
• Reposition the membrane in
the valve body (2)
• Integrate the flow sensor (3)
at the valve body (2)
(1)
(2)
(3)
(3)
www.device.airliquidehealthcare.com
YN033552
- Juin 2018 version
Note: pressing the padlock locks the
touchscreen.
To unlock = press the padlock and confirm