Errors and Warning Messages
15. Errors and Warning Messages
Info
Warning
The surface The sample temperature is lower than the
temperature is under .C sample temperature alarm.
Warning
The surface The sample temperature is higher than the
temperature is over ..°C sample temperature alarm.
Single measurements can You can delete measurements only within the
no longer be deleted. active checkzone.
Calibration interval expired: Appears when the instrument is switched on
Effect, Color, Color reference
and the calibration interval has expired.
Perform calibration.
Error
Memory full The maximum number of measurements has
been stored. Delete data which is no longer
needed.
Pins not pressed! Appears when sensor pins are not
pressed during a reading. Confirm and repeat
the reading.
Wrong Standard The wrong standard was used for black
Use Black Trap calibration.
Wrong Standard The wrong standard was used for white
Use White Standard calibration.