3 | Getting Started BYK-Gardner GmbH
24 spectro2guide Family 295 024 855 E 2405
Gloss (*)
Turn gloss measurement on or off.
Always use autostandard
Search for nearest standard is always active.
Geometry
Select d:8 measurement condition. Default is
Spin. Available for cat.no. 7070, 7086. Details
see Geometry [}51].
Always continue last test series
No new test series is created; latest series is
opened automatically.
Use gloss for movement detection (*)
An error message appears if instrument is moved during measurement. Default is activated. If
message appears without movements, deactivate this option – e.g. when measuring very thin
foils.
(*) Gloss measurement not included in spectro2go XS due to small aperture.
Additional Parameter
Depending on the model of your instrument additional measurement parameter can be avail-
able.
Fluorescence
Turn fluorescence indices on or off. Available for spectro2guide and spectro2guide Pro. De-
tails see Fluorescence Measurement [}36].
Jetness indices
Turn jetness and grayness indices on or off. Available for spectro2guide Pro. Details see Jet-
ness Measurement [}37] and Grayness Measurement [}39].
Setting Measurement Parameter
To set measurement parameter:
• Click on the parameter you want to set.
• A respective list with selectable parameters appears.
• Choose the required parameter and confirm via the checkmark in the upper right corner.
• Repeat these steps for all desired parameters.
Confirm by clicking on the checkmark in the upper right corner. More details see Measurement
Parameter [}47].
NOTICE
• Changes in the measurement parameter apply to new standards only.
• Standards with existing measurement data will be measured with previous parameter to en-
sure data consistency.