3 | Getting Started BYK-Gardner GmbH
24 spectro2guide Family 295 024 855 E 2405
Gloss (*)
Turn gloss measurement on or off.
Always use autostandard
Search for nearest standard is always active.
Geometry
Select d:8 measurement condition. Default is
Spin. Available for cat.no. 7070, 7086. Details
see Geometry [}51].
Always continue last test series
No new test series is created; latest series is
opened automatically.
Use gloss for movement detection (*)
An error message appears if instrument is moved during measurement. Default is activated. If
message appears without movements, deactivate this option – e.g. when measuring very thin
foils.
(*) Gloss measurement not included in spectro2go XS due to small aperture.
Additional Parameter
Depending on the model of your instrument additional measurement parameter can be avail-
able.
Fluorescence
Turn fluorescence indices on or off. Available for spectro2guide and spectro2guide Pro. De-
tails see Fluorescence Measurement [}36].
Jetness indices
Turn jetness and grayness indices on or off. Available for spectro2guide Pro. Details see Jet-
ness Measurement [}37] and Grayness Measurement [}39].
Setting Measurement Parameter
To set measurement parameter:
Click on the parameter you want to set.
A respective list with selectable parameters appears.
Choose the required parameter and confirm via the checkmark in the upper right corner.
Repeat these steps for all desired parameters.
Confirm by clicking on the checkmark in the upper right corner. More details see Measurement
Parameter [}47].
NOTICE
Changes in the measurement parameter apply to new standards only.
Standards with existing measurement data will be measured with previous parameter to en-
sure data consistency.