6 | System Configuration BYK-Gardner GmbH
58 spectro2guide Family 295 024 855 E 2405
6.2.4 Disconnect Network
To disconnect from the network, switch the WiFi option off in the WiFi configuration screen.
The Wifi icon disappears in the main menu.
6.3 Instrument Calibration
The system delivery comprises a calibration tile and test tiles for color and gloss, see Test and
Calibration Standards [}16]. These tiles can be used to check if your instrument is measuring
correctly. Each tile comes with a dedicated serial number matching your instrument. In the con-
figuration menu select Calibration.
Information
Check date of last calibration, serial numbers
and values for calibration and test tiles here.
White standard
Comprises the options for instrument calibra-
tion.
Autodiagnosis
Comprises the options for auto diagnosis.
Calibrate on white standard
Start the manual calibration procedure here.
Docking station interval
Select interval for the auto diagnosis here.
Possible values: 1, 3, 7, 14 days. Default: 7
days.
Add standard to list
Add calibration tile to the list of standards in
the instrument.