CALIBRATION
REQUIRED
3837
| Operating Instructions Manual | Operating Instructions Manual
CALIBRATION REQUIRED ERROR
This pop-up will remain in place regardless of whether the error automatically corrects itself. You will be
required to turn the unit off and then on again to attempt to clear the error. When the error occurs, the
system will go from the Firing or Ready state to the Standby state. No device settings or controls shall be
available. You should send the laser for calibration to AMD Lasers or nearest Authorized Service Center
as soon as possible.
ELECTROMAGNETIC COMPATIBILITY (EMC)
The Picasso Plus and Picasso Lite Plus Soft Tissue Diode Lasers need special precautions regarding
EMC and need to be installed and put into service according to the EMC information provided in this
manual.
Portable and mobile RF communications equipment can affect the Picasso Plus or Picasso Lite Plus. The
use of accessories and cables other than those specified by the manufacturer, AMD Lasers may result in
increased emissions or decreased immunity of the Picasso Plus or Picasso Lite Plus. The Picasso Plus or
Picasso Lite Plus should not be used adjacent to or stacked with other equipment and that if adjacent or
stacked use is necessary, the Picasso Plus or Picasso Lite Plus should be observed to verify normal
operation in the configuration in which it will be used.
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The Picasso® Plus and Picasso Lite® Plus are intended for use in the electromagnetic environment specified below. The
customer or the user of the Picasso Plus or Picasso Lite Plus laser should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The Picasso Plus and Picasso Lite Plus uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Not available
Voltage Fluctuations / Flicker
emissions
Not available
The Picasso Plus and Picasso Lite Plus are suitable for use in all establishments other than domestic and those directly
connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.