Maintenance | 6-17
System Status. 0=Everythingok,Not
0=something is wrong
SystemState.100=NotReady,201=Measure
Run,502=Idle,600=diag
CurrentOutput1Parameter
CurrentOutput1OutputType.0=4-20 mode,
1=0-20 mode.
Ifoverangeenable(ANAOUT1OR=1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Faultoption(ANAOUT1F)
CurrentOutput1OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput1FaultOption.0=Nofault,
1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT1T=0):
Lowendofrangebecomes3mA
Outputratiobecomes 0.
CurrentOutput2Offset-
CurrentOutput2Parameter-
CurrentOutput2OutputType.0=4-20 mode,
1=0-20 mode.
Ifoverangeenable(ANAOUT2OR=1)
0 = 3.78 - 20.5, 1 = 0-20.5, high or low
values can also be set by fault value, See
Faultoption(ANAOUT2F)-
CurrentOutput2OverrangeEnable.
0=disabled, 1=enabled
CurrentOutput2FaultOption.0=Nofault,
1=Lowfault,2=Highfault.
Ifhighfault(2):
Highendofrangebecomes22mA
Outputratiobecomes1
Iflowfault(1)ANDin4-20 mode
(ANAOUT2T=0):
Lowendofrangebecomes3mA
Outputratiobecomes0.
Relay1.0=OPEN,1=CLOSED
Relay1Delay.Holdstherelayoutputforthis
set amount of time.
Relay1Function.0=Disabled,
1=Concentration, 3=Data Valid, 4=System
Relay1Source.Thesourceparameter
number used to evaluated the relay function.
Relay1Value.Thevalueusedtoevaluate
the relay function.
Relay2.0=OPEN,1=CLOSED
Relay2Delay.Holdstherelayoutputforthis
set amount of time.
Relay2Function.0=Disabled,
1=Concentration, 3=Data Valid, 4=System
Relay2Source.Thesourceparameter
number used to evaluated the relay function.
Relay2Value.Thevalueusedtoevaluate
the relay function.
Relay3.0=OPEN,1=CLOSED
Relay3Delay.Holdstherelayoutputforthis
set amount of time.
Relay3Function.0=Disabled,
1=Concentration, 3=Data Valid, 4=System
Relay3Source.Thesourceparameter
number used to evaluated the relay function.
Relay3Value.Thevalueusedtoevaluate
the relay function.
Relay4.0=OPEN,1=CLOSED
Relay4Delay.Holdstherelayoutputforthis
set amount of time.
Relay4Function.0=Disabled,
1=Concentration, 3=Data Valid, 4=System
alarm
Relay4Source.Thesourceparameter
number used to evaluated the relay function.
Relay4Value.Thevalueusedtoevaluate
the relay function.
Master enable for all relays. 0=Disabled,
1=Enabled
Concentration Unit of Measure for analyte 2
Decimal places to use for displaying analyte 1
Decimal places to use for displaying analyte 2
Normally closed indicator for relay 1.
1=Normally closed, 0=Normally open
Normally closed indicator for relay 2.
1=Normally closed, 0=Normally open
Normally closed indicator for relay 3.
1=Normally closed, 0=Normally open
Normally closed indicator for relay 4.
1=Normally closed, 0=Normally open
Temporary cache to miror warnings in
SYSALARM1
Temporary cache to miror warnings in
SYSALARM2
ConcentrationofReferencecall
Analogoutput1,0- follow source during cal,
1 - hold output during cal
Analogoutput2,0- follow source during cal,
1 - holdoutputduringcalL82