2-4 | Model 5100 HD Analyzer
Sample Temperature
-20C to +150C (-4 F to +302 F) Gas-phase samples
only
Ambient temperature -20
o
C to +50
o
C (-4
o
F to +122
o
F)
Storage temperature -25 C to +80 C (-13 F to +176 F)
Ambient humidity 0-90%, non-condensing
Physical
880 mm x 674 mm x 302 mm (34.6” x 26.5” x 11.9”),
60 kgs (132 lbs)
Enclosure material 300 series stainless steel
Maximum altitude 2000 meters
Pollution Degree 2
Ingress protection (IP) IP65 and NEMA Type 4X
Sample ow rate 1-10 lpm (2-20 SCFH)
Sample gas Natural gas, Hydrogen, Carbon Dioxide, Propane,
Oxygen, and others pending application review
Sample wetted materials
Stainless steel, SiO2 glass, Gold, Teon, EPDM
rubber
Electrical Classications This analyzer will be certied with regard to
electrical safety for Pollution Degree 2 (normally
non-conductive environmental pollution occurs
with occasional condensation) and Installation
Category II (local level transients, less than
those found at power distribution levels). Mains
supply voltage uctuations are not to exceed 10
percent of the nominal supply voltage. Other ap-
provals will include acceptance for use in
NEC/CEC Class I Division 2 Groups A, B, C, D T3
and world-wide group II Zone 2 IIC T3 classied
hazardous areas.