AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 37 / 60
Current Measuring on Channel 2:
Measuring Instrument (AC current)
1. Instrument Scope model
2. Instrument value Voltage rms
Measuring Equipment
3. Equipment Type Current Probe
4. Equipment Agilent N2783A
B: Current Measuring Instrument:
The current measuring is realized with a current probe connected to a measurement receiver. This instrument will
measure the transferred ac voltage. The selected AC current instrument must be able to measure up to 100kHz
LeCroy preferred manufacturer by EM Test
Connect Instrument
Press Connect for check the instrument connection.
Setup : for add or modify an instrument in the library
C: Measuring Equipment:
The current sensor is a current transformer or a shunt resistor probe connected to a measurement receiver. This
instrument will measure the transferred ac voltage. The selected AC current instrument must be able to measure
up to 100kHz.
Transform the current into a rms ac voltage
12.4.2 General Information about Magnetic Immunity
Below is an overview of the field strength limits using various AMP 200N models.