AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 42 / 60
Test characteristics for CI 210
Method:
Coupling Method : Conducted
Test Regulation Method: Voltage Level ( Closed Loop)
Disturbance Current limit : yes/no setted value
Diagram Unit X-Axis : kHz ; Log
Y-Axis: Vpp ; linear
Example Closed Loop
Tolerances
Tolerances : positive +10%
negative -0%
Additional Settings (Closed Loop with Scope only)
Level Change at next Testpoint: 0dB
Couplings
Coupling Type none (direct)
Coupling Device none
User Software settings (Closed Loop with Scope only)
The user has to complete the test settings with his individual used measuring devices.
New measuring instruments can be installed through the Setup Menu in Mainwindow of Auoewave.control or by
clicking the setup button in the Voltage Measuring panel by the user.
A: Voltage Measuring Instrument: CH1
Measurement Instrument
Measuring AC voltage ( pk-pk)
Scope
Mandatory in Ford EMC-CS 2009 standard
Instrument Value
Voltage Peak To Peak
B: Measuring Equipment (voltage probe):
Equipment Type:
Voltage Probe
Equipment:
LeCroy HFP 2500 10:1
Remark
The 200mV pk-pk voltage ripple signal at the test start needs a signal of only 33mVpk-pk in the
signal generator (range ± 10V). Additional the VDS200 noise is specified as < 0.2V pk-pk.
Therefore low signals will have a noise ripple who loock very high.