AMETEK CTS AMP 200N series
Manual for Operation V 4.0.3 56 / 60
Test characteristics Default settings as per Standard
Method:
Coupling Method : Radiated
Test Regulation Method: Verify H-Field
Diagram Unit X-Axis : kHz ; Log
Y-Axis: dBpT ; Log
Tolerances : +2.7% -0%
Additional Settings:
Start at Level : -20 dB
Level Change at next Testpoint: 0dB
If the first measurement is already too high the user can reduce the start Level ( example to -30dB)
User Software settings for adapt the hardware
The user has to complete the test settings with his individual used coupling and measuring devices. New
coupling devices can be installed in the menu Setup/Coupling Device by the user.
Coupling
Coupling Type Antenna
Coupling Device Radiating Loop
Select the antenna you are using for the test.
Setup for add or modify new Antennas in the Library in the Setup Coupling Devices
For more information refer to the Autowave software manual.