13.2 CI 250 Pulse verification
CI 250 Test Parameter requirements
Peak measurement
Un
Upeak
U1
5.00
U2
3.65
U3
2.67
U4
1.95
U5
1.42
U6
1.04
U7
0.76
U8
0.55
U9
0.41
U10
0.30
Tolerances:
Time interval ± 10%
Voltage +10% -0%
Climatic Test conditions
Temperature 23± 5.0 degrees C
Humidity 20 to 80% relative humidity (RH)