4) Turnthescanneraround180°.YoushouldnowbelookingatthesidewithaPCB
muchclosertothefrontedgeofthecasing.LocatethepinsmarkedLK3,again
towardstherighthandsideofthePCB.Notethatthesepinsarelocatedonthe
exposedpartofthesecondlayerofPCB.Replacethejumperremovedinstep3
ontheTOPTWOofthethreepinsofLK3marked‘InternalEmissivity’.
5) Thescannersemissivitycannowbesetusingthetowidenticalrotaryswitches
thatcanbeseentowardsthemiddleofthePCB.
Theleftswitchsetstherstdecimalplace,therightswitchsetsthesecond
decimalplace.Forexamples,seethetablebelow.
Left switch setting Right switch setting Emissivity setting
8 5 0.85
7 2 0.72
0 0 1.00