Page 3
User Guide LSPHD Linescanner Head
1.2 Specications-LSPHD Series
1.2.1 Common specications
Ambient temperature range 5to60°C/41to140°F(inspecication)
5to70°C/41to158°F(operating)
Alignment Class 2 Laser indicating scan plane and extent
Laser scan rate 1 Hz
Internal Ambient Temperature Sensor
(IATS)
-10to90°C/14to194°F
Emissivityrange
(remoteorinternallyset)
0.20 to 1.00
Sealing IP65
EMC
LVD
EN61326ClassA[emissions&immunity]
EN61010-1
Vibration(Anyaxis) 2g[10to30Hz]
3g[30to300Hz]
1.2.2 LSPHD 1 series specications
LSPHD 10 LSPHD 11
Operatingwavelength 1µm 1µm
Temperature range 600 to 1400 °C
1112 to 2552 °F
700to1500°C
1292to2732°F
SystemAccuracy
(1)
(5to95%ofrange)
±2°C/3.6°F
Focus distance
(2)
1m/39.7intoinnity(continuously
adjustable)
FieldofView
(staticto95%ofradiance)
500:1(0.002radians)
Repeatability <0.5°C/0.9°F
Temperature Resolution
(3)
< 5% of range
(noise) < 15% of range
< 30% of range
Above 50% of range
≤6°C/10.8°F
≤6°C/10.8°F
≤4°C/7.2°F
≤1°C/1.8°F
Scan Angle
(4)
80°
Scan Rate Factoryset
Useradjustable
50 Hz
10 to 150 Hz in 1 Hz steps
Speed of response
(Tau-linearoutputonrisingedge)
≤1µs
AmbientStability
deltaT[indicated]/deltaT[ambient]
≤1°/10°
(1)Includeserrorcorrection
(2)Focusisfactorysetto2m
(3)4lineaveragingbelow15%ofrange
(4)Scananglemaybesetintherange40to80°bythePCsoftware