16308-101 Rev. A • 29
Table 202 – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity
All Medical Electrical Equipment and Medical Electrical Systems
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
Tono-Vera Vet is suitable for use in electromagnetic environment specified below. The customer or user of Tono-Vera Vet should ensure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE LEVEL ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Electrostatic
Discharge
IEC 61000-4-2
±8kV Contact
±2kV, ±4kV, ±8kV, ±15kV Air
±8kV Contact
±2kV, ±4kV, ±8kV, ±15kV Air
Floors should be wood, concrete or ceramic tile. If floors are synthetic, the R/H should be at
least 30%.
Electrical Fast
Transients/Bursts
IEC 61000-4-4
±2kV Mains Power Lines
±1kV I/O Lines
100 kHz repetition frequency
±2kV Mains Power Lines
±1kV I/O Lines
100 kHz repetition frequency
Mains power quality should be that of a typical residential, commercial or hospital
environment.
Surges
IEC 61000-4-5
±0.5kV, ±1kV
Line-to-line
±0.5kV, ±1kV, ±2kV
Line-to-ground
±0.5kV, ±1kV
Dierential Mode
±0.5kV, ±1kV, ±2kV
Common Mode
Mains power quality should be that of a typical residential, commercial or hospital
environment.
Voltage Dips
IEC 61000-4-11
0% Ut, 0.5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270°, 315°
0% Ut, 0.5 cycle at 0°, 45°, 90°,
135°, 180°, 225°, 270°, 315°
Mains power quality should be that of a typical residential, commercial or hospital
environment. If the user of Tono-Vera Vet requires continued operation during power mains
interruptions, it is recommended that Tono-Vera Vet be powered from an uninterruptible power
supply or battery.
0% Ut, 1.0 cycle and
70% Ut, 25/30 cycles
Single phase: at 0°
0% Ut, 1.0 cycle and
70% Ut, 25/30 cycles
Single phase: at 0°
Voltage
Interruptions
IEC 61000-4-11
0% Ut, 250/300 cycles 0% Ut, 250/300 cycles
Power Frequency
50/60Hz
Magnetic Field
IEC 61000-4-8
30A/m
50 Hz or 60 Hz
30A/m
50 Hz or 60 Hz
Power frequency magnetic fields should be that of a typical residential, commercial or hospital
environment.
Guidance Tables (continued)