Sequoia / Tahoe Series User Manual California Instruments
184 M447352-01 REV-A
The top-level menu of the CONFIGURATION screen is shown in Figure 5-141. It can be reached in one
of two ways:
1. Tapping CONFIGURATION on the HOME Screen of the front panel touch-screen.
2. Scrolling to CONFIGURATION with the encoder and depressing the encoder switch.
The UP arrow button will return to the previously selected screen menu (in this case the HOME Screen).
The HOME button will return to the HOME screen that has the top-level menu for the sub-menu being
displayed; for the CONFIGURATION screen top-level menu, is the HOME Screen.
Figure 5-141. CONFIGURATION Screen Top-Level Menu
The following sub-menus are available in the CONFIGURATION menu:
Entry Description
SYNC SETTINGS Programs the Sync Voltage, Sync Phase of each output phase,
and Sync Freq.
Figure 5-142. Sync Settings
VOLTAGE PROTECTION Programs the Voltage Protection threshold for the output
voltage of each output phase. Exceeding the OVP threshold will
result in the shutdown of the output, with the output isolation relay
opened and the output voltage programmed to zero. The
maximum OVP setpoint is 115% of full-scale low-range/high-
range output voltage (191V/383V in theAC and (AC+DC)-modes,
and 253V/506V in the DC-mode, . The default value is 115% of
full-scale.