EasyManua.ls Logo

Ametek Sorensen DCS8-350E - Remote Monitoring of Readback Signals and Status Indicators; Readback Signals; Status Indicators

Ametek Sorensen DCS8-350E
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Sorensen DCS-E Series 3kW Supplies Advanced Operation
M362295-01 Rev D 4-21
4.6 REMOTE MONITORING OF READBACK
SIGNALS AND STATUS INDICATORS
4.6.1 Readback Signals
Calibrated readback signals for remote monitoring of the output voltage and current are
available via connections at the J3 connector on the rear panel. Internal switch SW1 settings
allow you to select either a 0-5 Vdc or a 0-10 Vdc range for the output. See Section 4.2
Configuring for Remote Programming, Sensing, and Monitoring for more information about
making these connections.
The following table shows the required pin connections and switch settings for remote
monitoring of readback signals with 0-5 Vdc or 0-10 Vdc outputs. Use shielded-twisted pair
wiring (20 to 24 AWG) and ground the shield to J3 connector pin 6 or to the chassis via one of
the J3 connector screws.
Readback Signal
J3 Connections: Switch SW1 Settings:
Output Signal
Range
Signal (+) Return (-) Switch # Setting
Output Voltage Pin 19 Pin 12 SW1-4 CLOSED 0-5 Vdc
OPEN 0-10 Vdc
Output Current Pin 7 Pin 12 SW1-3 CLOSED 0-5 Vdc
OPEN 0-10 Vdc
The readback signal represents 0 to 100% of the model-rated output.
4.6.2 Status Indicators
Status indicators for thermal shutdown, OVP circuit, programming mode, and operating mode
are available via connections on the J3 connector on the rear panel.
The following table shows the indicator signals, the J3 connector pin at which they are
available, an approximation of the signal magnitude, and the source impedance through which
the signal is fed. Use 20 to 24 AWG wiring.
Indicator Signal/
Alternate State
J3 Connections:
Signal
Voltage
Source
Impedance
Signal (+) Return (-)
Thermal Shutdown/ Pin 18 Pin 6 +13V 1k ohms
Normal Operation Pin 18 Pin 6 0V 1k ohms
OVP Circuit Activated/ Pin 17 Pin 6 +13V 1k ohms
OVP Circuit Not Activated Pin 17 Pin 6 -13V 1k ohms
Remote Programming Mode/ Pin 4 Pin 6 +13V 1k ohms
Local Programming Mode Pin 4 Pin 6 0V 1k ohm
Voltage Mode Operation/ Pin 5 Pin 6 +13V 1k ohms
Current Mode Operation Pin 5 Pin 6 -13V 1k ohms

Table of Contents