41
Step/Range
Settings
Adjust for:-
Load
Detail
17, Aux o/p
5V
5.00V ± 10mV
DVM
o/p V span
18, Aux o/p
5V
5.00V on display
DVM
V readback
19, Aux o/p
1V
1.00V on display
DVM
V readback
20, Aux o/p
5V
5.00V ± 0.5mV
DVM
V readback
21, Aux o/p
off
0.000 flashing 0.001
A readback zero
22, Aux o/p 5V Display equals load setting
or ammeter
2.7A load, or
1.8 in series
with ammeter
A readback
XDL 56-4, XDL 56-4P, XDL 56-4T & XDL56-4TP
Step/Range
Settings
Adjust for:-
Load
Detail
1, 56V 2A
0.01V 2A
10mV ± 0.5mV
DVM
o/p 0V
2, 56V 2A
0.01V 2A
0.010V on display
DVM
V readback zero
3, 56V 2A
35V 2A
35V ± 0.5mV
DVM
o/p V span
4, 56V 2A
35V 2A
35.000 on display
DVM
V readback span
5, 56V 2A
41V 2A
41.000 on display
DVM
V readback 13bit
6, 56V 2A
0.01V 2A
10mV ± 0.5mV
DVM
o/p 0V
7, 56V 2A
35V 2A
35V ± 0.5mV
DVM
o/p V span
8, 56V 2A
2V 0.001A
1mA ± 0.5mA
milli-ammeter
o/p A offset
9, 56V 2A
2V 0.001A
0.001 on display
milli-ammeter
A readback zero
10, 56V 500mA
2V 0.1mA
0.1mA ± 0.05mA
milli-ammeter
o/p mA offset
11, 56V 500mA
2V 0.1mA
flashing 0.0/0.1 on display
milli-ammeter
mA readback zero
change load
12, 25V 4A
2V 4A
4A ± 0.5mA
ammeter
o/p A span
13, 25V 4A
2V 4A
4.000 on display
ammeter
A readback span
14, 56V 500mA
2V 400mA
400mA ± 0.05mA
ammeter
o/p mA span
15, 56V 500mA
2V 400mA
400.0 on display
ammeter
mA readback span
16, 56V 500mA 2V 410mA 410.0 on display ammeter mA readback 13bit
17, Aux o/p
5V
5.00V ± 10mV
DVM
o/p V span
18, Aux o/p
5V
5.00V on display
DVM
V readback
19, Aux o/p
1V
1.00V on display
DVM
V readback
20, Aux o/p
5V
5.00V ± 0.5mV
DVM
V readback
21, Aux o/p
off
0.000 flashing 0.001
A readback zero
22, Aux o/p 5V Display equals load setting
or ammeter
2.7A load, or
1.8 in series
with ammeter
A readback