Remote Operation
5-20 M370430-01 Rev B
Multiple Power Supply Setup
Master Setup:
Configure the master XG/XTR by selecting the communication
interface you wish to use to communication with the Master and
follow the setup instruction in this chapter.
Important: If either RS-232 or RS-485 are used for communication with the
master, the data rate must be configured for 9600 bps to properly communicate
with the slave units.
Slave Setup:
1. Turn the 9-position mode control to PGM.
is displayed in the output voltage display.
2. Turn the rotary Adjust/Enter control to select and press
the control to enter the setting.
3. Set a unique address. See “Multichannel Address Setting” on
page 5-25.