Conguration & Calibration
25
3.6 Description of Function Keys (Continued)
Figure 3-3
PC Software Menu Screen
automatically communicates
all "name plate data" from
transmitter
Write to Transmitter [F5 on keyboard]
Sends new or edited conguration data to the transmitter.
Data elds that have been edited but not sent to the
transmitter are displayed in red.
Real Time View [F4 on keyboard]
Displays the real time values of level, capacity, distance,
temperature, loop current, percentage, and status.
Point Calibration [F6 on keyboard]
Calibrates the HART® protocol transmitter using Point
calibration. See Section 3.8 Calibration. Enter the low
point and high point of level for an accurate calibration.
D/A Trim
Allows a eld reference meter to be connected to the
transmitter for adjusting transmitter output current.
See Section 3.10.
Strapping Table
Displays the values of the input to level and output to
volume in percent in a 21-point table. Allows points to
be changed to accommodate irregularly shaped vessels.
See Section 3.11.
Congure Meter
Congures the optional Digital Integral Meter (440-44-3)
used for local indication. See Section 3.12.
True Level Calibration (grayed out)
This button is inactive for the Universal III Transmitter.
It is for the extra features that come with the True Level
model transmitter.