Conguration & Calibration
27
Figure 3-6
Vessel Conguration from
Menu screen
Figure 3-7
LRV & URV Conguration
from Menu screen
3.7.2 Vessel Configuration
A. Select Vessel Units. The default is gallons. Press Enter
and choose the units that correspond to the vessel
measurement. The units include both weight and
volume outputs. Press Tab or Enter on your key board
to continue.
B. Edit the
Maximum Capacity of the vessel. Enter the
corresponding value of weight or volume equal to the
Maximum Level. Enter 100 for percent if the weight
or volume units are not known or needed. Press Tab or
Enter on your keyboard to continue.
C. Select
Vessel Type. Available options include:
• Vertical Tank (V
ertical)
• Horizontal cylinder with at ends (H
rzcy/Fl)
• Horizontal cylinder with dished ends (
HrzCy/Ds)
• Horizontal cylinder with hemispherical ends
(Hrzcy/Hm)
• Spherical (S
pHere)
• The default is
Vertical. Press Enter and choose the
type of vessel.
D. Click on Write to Transmitter.
E. Move on to Range Values (
URV & LRV) section of
menu.
3.7.3 Lower and Upper Range Values (LRV and URV)
Enter the LRV and URV to set the current (mA) window of
the vessel.
A. Edit
LRV (Lower Range Value) to display the output
you want to see when the transmitter generates 4 mA
current. The default
LRV is 0 feet.
B. Edit
URV (Upper Range Value) to display the output
you want to see when the transmitter generates 20 mA
current. The default
URV is 100 feet for the Universal
III.
C. Click on
Write to Transmitter. Conguration is now
complete.