Conguration & Calibration
29
Table 3-1
Span Range Position for a number of common Sensing Elements
Jumper Position = 1 2 3 4 5 6
Maximum pF = 20 100 450 2000 10000 45000
SYSTEM # SENSOR # MAXIMUM PROBE LENGTH IN FEET
CONDUCTING LIQUIDS:
509-75-X09 700-1-22 N/A N/A 5.4 20 N/A N/A
509-75-X25 700-5-54 N/A N/A 9.2 40 200 920
509-75-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-75-X13 700-5-18 N/A N/A 1.4 6.25 31 140
509-75-X07 700-2-24 N/A N/A 7.8 20 N/A N/A
509-75-X05 700-2-37 N/A N/A N/A N/A 4.1 14
509-77-X06 700-1-24 N/A N/A 5.4 20 N/A N/A
509-75-X30 700-1-62 N/A N/A 5.4 20 N/A N/A
509-75-724 700-5-29 N/A N/A N/A 4.2 21 95
INTERFACE APPLICATION:
509-76-X06 700-2-57 N/A N/A 1.2 5.3 20 N/A
509-76-X02 700-2-27 N/A N/A 1.0 4.8 20 N/A
509-76-X04 700-2-37 N/A N/A N/A N/A 4.0 14
INSULATING K = 1.5-5:
509-77-X06 700-1-24 N/A 2.7 12.5 20 N/A N/A
509-77-X06 700-1-22 N/A 8.3 20 N/A N/A N/A
509-77-X06 700-5-54 N/A 8.0 37.5 166 N/A N/A
509-77-X06 700-2-57 N/A 3.6 16.2 20 N/A N/A
509-77-X06 700-5-18 N/A 3.8 17.1 76 N/A N/A
509-77-X06 700-2-24 N/A 6.0 20 N/A N/A N/A
509-77-X06 700-2-27 N/A 5.0 20 N/A N/A N/A
509-77-X06 700-2-37 N/A 5.0 14 N/A N/A N/A
Universal III Span Range Setting Chart
Probe Length vs. Span Position Number / maximum pF