Maximum Waveform LengthDynamic (Repeat) Looping, up to 32,767 times, and 512k data points
DCdata compression for waveforms exceeding (2048k, B version)
output memory length
Minimum Waveform Length8 Data points on Version 25 64 Data points
16 Data points on Version 100
Waveform Equation Memory30 kbytes non-volatile, expandable to 78k 78 kbytes non-volatile
(Stores waveform 1280 characters maximum file length 1280 characters maximum file length
descriptions by filename)
Waveform Equation Memory Retention: >2 years with supplied lithium batteries and a 50% power-on duty cycle
Memory Retention (provision is made for field replacement with more conventional Alkaline batteries)
AUXILIARY INPUTS/OUTPUTS (Menu driven setup)
External Clock Input DCto max. update rate of version DCto 800 MHz update rate
External Trigger Input Trigger Modes: Free Run, Manual, Triggered Start, Trigger Modes: Free Run, Manual, Triggered Start,
Triggered Stop, Gated by Trigger, and Triggered Triggered Stop, Gated by Trigger, and Triggered
Start/Stop Start/Stop; Delay: < (40 ns + 2 clock periods); Jitter:
Delay: < 100 ns +1 clock period < (100 ps + 1 clock period).
Jitter: < 10 ns
Trigger and External Clock Threshold: 0V or 1V; Slope: plus or minus; Threshold: 0V or 1.5V; Slope: plus or minus;
Input Parameters Impedance: 1 M|| 25 pF; Sensitivity: ±500mV Impedance: 10 k|| 50 pF; Sensitivity: ±500 mV
about selected threshold; Maximum input: ±30V; about selected threshold; Maximum input: ±30V;
Marker Output Negative transition at waveform start, positive Positive pulse at user selected Mark Time or Mark
transition at Mark Time or Mark Count (TTL, Count from start of waveform; Pulse duration:
10 unit loads) 32 data point clocks (TTL, drives a grounded 50load)
Sync Output Positive pulse at waveform start (TTL, 10 unit loads) Positive pulse at waveform start; Pulse duration: 32 data
point clocksTTL, drives a grounded 50load)
Multi-Unit Synchronization Via phase locked internal clocks and ganged trigger Via auxiliary inputs and outputs
functions in a Master/Slave configuration, up to 8
units, with rear panel adjustments for cable delays
Digital Waveform Output 16 bits plus clock, simultaneous with Analog Output;
TTL on Version 25; ECL (diff.) on Version 100 (opt.)
REMOTE INSTRUMENT CONTROL
Interface Optional IEEE-488 and/or RS-232 Standard IEEE-488 and RS-232
Transfer Data Rates IEEE-488: 2 kbytes/s standard IEEE-488: 300 kbytes/s using High Speed DMA
300 kbytes/s with High Speed DMA opt. RS-232: up to 9600 baud
RS-232: up to 9600 baud
WAVEFORM GENERATION MODES
Mathematical By Front Panel entry or by remote communications
(Polynomial) Notation
Standard Functions By Front Panel entry or by remote communications: Standard Functions include sine wave,
square wave, pulse, triangle, sawtooth, and random noise
Internal Noise Source Pseudo-random white noise with a normal Random noise with approx. normal amplitude
amplitude distribution; Cycle time: 168 seconds distribution can be added to waveforms in memory,
Bandwidth: 0.006 Hz to selectable 20 kHz, 200 kHz, with selectable noise amplitude, bandwidth, seed
or 2 MHz; Maximum Amplitude (RMS): 2.7V for 2 MHz(starting point), and signal+noise amplitude. Signal
BW, 0.95V for 200 kHz BW, 0.31V for 20 kHz BW may be repeated in memory so different noise
Amplitude range: 60 dB in 1 dB steps values occur over each signal cycle.
Direct Waveform Data Via IEEE-488 or RS-232
Download
Captured Data ReproductionWith Analogic Model 6100 or 6500 Waveform Analyzer (except Model 2000)
Scope Draw with Cursor For point by point waveform editing
Control
GENERAL
Specification Conditions 0 to 40°C operating temperature; 50resistive termination on both analog outputs, unless specified otherwise;
Open circuit output will double (±5%) in amplitude, but with some degradation in high frequency/speed perfor-
mance
Power Requirements 90-130 VACor 180-260 VACat 48-62 Hz; 90-130 VACor 180-260 VACat 48-62 Hz; 200W max
100W max
Dimensions 5.5"H x 16.5"W x 17"D (13.97 x 41.91 x 43.18 cm) 5.5"H x 16.5"W x 17"D (13.97 x 41.91 x 43.18 cm)
Weight 25 lb (11.3 kg) 25 lb (11.3 kg)