Command Directory 3-4 Commands
MG369xC GPIB PM PN: 10370-10374 Rev. F 3-21
CM5 or CM5 <value> <unit>
Description: Sets CW mode at M5 and opens the M5 parameter.
Places the MG369xC in CW mode at the preset (or previously set) M5 frequency and
causes M5 to be output. Also permits setting the M5 parameter. The frequency setting is
determined by the frequency range of the particular MG369xC model.
Functional
Category: CW FREQUENCY
Value: Frequency range of the MG369xC model
Units: GH, MH, KH, HZ
Output Command: OM5
Related
Commands: M5, ACW
Front Panel Key: Frequency | CW | Frequency Control > | Frequency List. . . | highlight M5, then press
Edit Selection or Output Frequency
CM6 or CM6 <value> <unit>
Description: Sets CW mode at M6 and opens the M6 parameter.
Places the MG369xC in CW mode at the preset (or previously set) M6 frequency and
causes M6 to be output. Also permits setting the M6 parameter. The frequency setting is
determined by the frequency range of the particular MG369xC model.
Functional
Category: CW FREQUENCY
Value: Frequency range of the MG369xC model
Units: GH, MH, KH, HZ
Output Command: OM6
Related
Commands: M6, ACW
Front Panel Key: Frequency | CW | Frequency Control > | Frequency List. . . | highlight M6, then press
Edit Selection or Output Frequency
CM7 or CM7 <value> <unit>
Description: Sets CW mode at M7 and opens the M7 parameter.
Places the MG369xC in CW mode at the preset (or previously set) M7 frequency and
causes M7 to be output. Also permits setting the M7 parameter. The frequency setting is
determined by the frequency range of the particular MG369xC model.
Functional
Category: CW FREQUENCY
Value: Frequency range of the MG369xC model
Units: GH, MH, KH, HZ
Output Command: OM7
Related
Commands: M7, ACW
Front Panel Key: Frequency | CW | Frequency Control > | Frequency List. . . | highlight M7, then press
Edit Selection or Output Frequency