Performance Verification 2 3-13 DVB-T/H Signal Analyzer (Options 64 and 57) Verification
MS2721B MM PN: 10580-00177 Rev. D 3-81
3. On the MG3700A, press the yellow Preset key located on the upper-left side of the instrument.
4. Press the Down Arrow key to select Yes.
5. Press the Set key (Note that two Set keys are available, and they both do the same thing).
6. Press the F1 soft key to select Load File to Memory.
7. Press the F1 soft key again to select Select Package.
8. Use the Down Arrow key to step through the selection list until the Digital_Broadcast option is
highlighted.
9. Press the Set key.
10. Press the F6 soft key, labeled Return.
11. Press the Set key.
12. Use the Down Arrow key to step through the selection list until the Digital_Broadcast option is
highlighted.
13. Press the Set key.
14. Use the Down Arrow key to step through the selection list until the DVB-T_H_00 option is highlighted.
15. Press the Set key.
16. Perform a Zero/Cal on Sensor A and Sensor B on the power meter. Set the calibration factor of both
sensors to 474 MHz.
17. Set the MG3700A Frequency to 474 MHz.
18. Set the Level to 25 dBm.
19. Ensure that the Mod On/Off key and the Output key each have an illuminated LED.
20. Turn on the power amplifier and allow it to warm up for at least 5 minutes.
21. Adjust the MN63A Attenuator so that the Sensor A reading is –10 dBm ± 1 dBm. Record the actual
attenuation reading in the “MS63A Attenuation Reading AT(–10)” column of Table A-130, “Readings for
Sensor A, Sensor B, DAB(–10), and MN63A Attenuation AT(–10), Option 64” on page A-41.
22. On the MG3700A, adjust the power level so that the Power Meter Sensor A reading
is 10.0 dBm ± 0.2 dBm.
23. Record the following values to the appropriate table cells in Table A-130:
Power Meter Sensor A reading to “Sensor A Reading”
Power Meter Sensor B reading to “Sensor B Reading”
24.
Subtract the value of the Sensor A reading from the value of the Sensor B reading and record the result to
the the “DAB (–10)” column of Table A-130.
25. Calculate the AT(set) values for Test Levels –10 dBm through –45 dBm and record the values in the
“AT(set) (dB)” column in Table A-133, “DVB-T/H Signal Analyzer Level Accuracy Measurement
Channel = 21ch at 474 MHz, Option 64” on page A-42.

Table of Contents

Other manuals for Anritsu Spectrum Master MS2721B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Anritsu Spectrum Master MS2721B and is the answer not in the manual?

Anritsu Spectrum Master MS2721B Specifications

General IconGeneral
BrandAnritsu
ModelSpectrum Master MS2721B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals