Performance Verification 1 2-2 Spectrum Analyzer Function Verification
MS2721B MM PN: 10580-00177 Rev. D 2-13
4. Set the Agilent 8648D Signal Generator RF to OFF ("RF OFF" appears on the right side of the display).
5. Connect the power sensor to the splitter "output" and adjust the power of the MG3692x so that ā20 dBm
appears at the splitter "output" (approximately +16 dBm on the MG3692x).
6. Turn off the RF of the MG3692x Synthesized Signal Generator.
7. On the Agilent 8648D Signal Generator, turn on the RF and adjust the RF amplitude so that ā20 dBm
appears at the splitter "output" (approximately +16 dBm on the 8684D).
8. Set the MS2721B to Spectrum Analyzer mode and then preset the instrument.
9. Set the MS2721B as follows:
Center Frequency: 600.151 MHz
Span: 100 Hz.
RBW: 30 Hz
VBW: 1 Hz
Reference Level: ā15 dBm
Attenuation Level: 0 dB
Pre Amp: Off
10. Press the Shift key and the Sweep (3) key. Press Detection, and press RMS (a red dot appears on the
label).