Performance Verification 1 2-2 Spectrum Analyzer Function Verification
MS2721B MM PN: 10580-00177 Rev. D 2-27
8. Adjust the MG3692x output power so that the power meter displays a reading that matches the Sensor B
reading for –30 dBm in Table A-21.
9. On the MS2721B, press the Amplitude function hard key, then set the Reference Level to –20 dBm.
10. Press the Freq function hard key, and press the Center Freq soft key.
11. Enter 10.1 MHz (or the next frequency).
12. Press the Amplitude function hard key, then set the Attenuation Level to 0 dB.
13. Press the Marker key, and press the Peak Search soft key.
14. Record the Marker 1 amplitude reading in Table A-22, “Amplitude Accuracy Across Frequency
Verification at 10.1 MHz” on page A-8.
15. Verify that the Marker 1 amplitude reading is within the specification.
16. Repeat Step 12 to Step 15 for Attenuation Levels of 5 dB, 10 dB, and 20 dB.
17. Adjust the MG3692x output power so that the power meter displays a reading that matches the Sensor B
reading in Table A-21 for –2 dBm.
18. On the MS2721B, press the Amplitude function hard key, then set the Reference Level to 10 dBm.
19. Repeat Step 12 to Step 15 for Attenuation Levels of 30 dB, 40 dB, 50 dB, and 60 dB.
20. Repeat Step 5 to Step 19 for all of the frequencies of Table A-23 through Table A-32.