46 APx515 B Series Audio Analyzer: Analog I/O Specifications
515
Non-Harmonic Content Typically <-110 dB when Fs 75 kHz, increasing to -55
dB at Fs =80 kHz
Phase Offset Range (Split Phase) –179.999 to +180.000 deg
DC Offset Range ±12.00 Vdc balanced;
±6.00 Vdc unbalanced
DC offset limits maximum ac signal
Residual DC Offset 0.25% of Vrms setting [0.09% of Vpp
setting] + 100 μV
Noise Characteristics
Shape White (<5 Hz to 86.4 kHz),
Pink (<10 Hz to 86.4 kHz),
IEC 60268-1 or BS EN 50332-1
Amplitude Range 0 to 45.2 Vpp, balanced;
0 to 22.6 Vpp, unbalanced
Amplitude calibration is approximate
IMD Test Signals (requires option AML)
SMPTE & MOD
Lower Frequency (LF) 40 Hz to 1.00 kHz LF tone must be 1/6 • HF tone.
SMPTE Upper Frequency (HF) 2.00 kHz to 60.00 kHz
MOD Upper Frequency (HF) 240 Hz to 60.00 kHz
Mix Ratio (LF:HF) 10:1, 4:1 or 1:1
Amplitude Range 0 to 45.2 Vpp, balanced;
0 to 22.6 Vpp, unbalanced.
Amplitude Accuracy ±0.06 dB [±0.70%]
Residual IMD
1,2,3
0.0025% [-92 dB], 4:1 mix ratio
DFD & CCIF
Difference Frequency (Fdiff) 80 Hz to 2.00 kHz F
mean
= (F1 + F2)/2.
Mean Frequency (Fmean) 250 Hz to 60.00 kHz F
diff
= |F2 - F1|
F
mean
must be 6 • F
diff
Amplitude Range 0 to 45.2 Vpp, balanced;
0 to 22.6 Vpp, unbalanced.
Characteristic Specifications Supplemental Information