EasyManua.ls Logo

AP B Series - Input and Common Mode Specs

AP B Series
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
48 APx515 B Series Audio Analyzer: Analog I/O Specifications
515
Maximum Rated Input
125 Vpk, 120 Vdc any input to ground;
0.5 Vpk for unbalanced bnc shields
Input Impedance
Balanced 100 k || 230 pF, each side to gnd
Unbalanced 100 k || 230 pF to bnc shield Electronically floating, 0.5 Vpk max;
bnc shield to ground 500 || 22nF
Input Termination
Selectable 600 ±1%, 1.5 W max. Termination automatically opens in the 80 V range.
Input Coupling
Selectable DC or AC Typically <0.5 μA bias current with DC coupling, typically
<0.03 dB roll-off at 20 Hz with AC coupling
Input Ranges
250 mVrms to 80 Vrms, 10 dB steps Maximum ac signal is 88 Vac unbal, 115 Vac bal, in the
80V range
Common Mode Rejection
4
Max common mode signal range:
250 mV, 800 mV, 2.5 V ranges 80 dB, 5 Hz to 5 kHz;
70 dB, 5 kHz to 20 kHz.
±6Vpk
8 V range 50 dB, 5 Hz to 20 kHz ±12 Vpk
25 V range 50 dB, 5 Hz to 20 kHz ±60 Vpk
80 V range 45 dB, 5 Hz to 20 kHz ±120 Vpk
Input Related Crosstalk
(–120 dB + 0.3 µV) to 20 kHz R
s
600
Level (Amplitude) Measurement
Range
Balanced or bridging input < 1 µV to 115 Vrms
Unbalanced input < 1 µV to 85 Vrms
Accuracy (1 kHz) ±0.05 dB [±0.60%]
Flatness (1 kHz ref, DC coupling)
10 Hz to 20 kHz ±0.010 dB Typically < 0.005 dB.
20 kHz to 50 kHz ±0.030 dB
50 kHz to 80 kHz ±0.10 dB
Characteristic Specifications Supplemental Information