EasyManua.ls Logo

ATP Instrumentation 1309 - Spectral and Accuracy Specifications; Environmental and Physical Specifications

Default Icon
15 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4
¾ Spectral Response: CIE Photopic. (CIE human eye
response curve).
¾ Spectral Accuracy: CIE Vλ function f1’6%
¾ Cosine Response: f2’ 2%
¾ Accuracy: ±5% rdg±10d.(<10,000Lux)
±10% rdg±10d.(>10,000Lux)
¾ Repeatability: ±3%
¾ Sampling Rate: 1.5 times/sec of analog bar-graph
indication;1.5times/sec of digital display.
¾ Photo Detector: One silicon photo diode and spectral
response filter.
¾ Operating temperature & Humidity :
0 to 40(32to 104) & 0% to 80% RH.
¾ Storage Temperature & Humidity :
-10 to 50(14 to 140) & 0% to 70% RH.
¾ Power Source: 1 piece 9V battery.
¾ Photo detector Lead Length: 150cm (approx.);
¾ Photo detector Dimensions: 115L×60W×20H(mm);
¾ Meter Dimensions: 170L×80W×40H;
¾ Weight: 390g.
¾ Accessories: Carry case, instruction manual, battery.