PROBLEM POSSIBLE CAUSE SOLUTION
Inaccurate
Weighing
or
Unnished
Calibraon
Dierent Areas of
the Plaorm Give
Dierent Weight
Readings
Plaorm or feet are not level;
Load cell broken;
Objects between Load Cell &
scale
Ensure the plaorm is level and
that all four feet are on a solid,
level surface; Clear the objects;
Replace the Load Cell
Big Tolerance with
Full Corner
Repeang Funcon
Malfuncon
The plaorm is not level;
Broken Load Cell; Aging
problem with the PCB;
Objects between Load Cell &
scale
Ensure the plaorm is level and
that all four feet are on a solid,
level surface; Clear the objects;
replace the Load Cell
Calibraon Can’t
Be Finished
The inner code at zero point
is too small or too large; Scale
is unstable; Objects between
Load Cell & scale; Broken
Load Cell
Recalibrate the scale; Replace PCB;
Clear the objects; Replace the Load
Cell
Funcon
Problem
Key Buon
Malfuncon
Broken Key buon; Aging
problem of the apparatus on
PCB
Check the buon, replace if
necessary; Replace PCB
Not Communicang
Smoothly
Communicaon wire broken;
Interface apparatus broken
Replace wire; Replace PCB
Power is supplied by an internal rechargeable 6V 4Ah rechargeable battery.
When “Lo.bAt” is displayed, the battery must be recharged.
Plug in the AC power adapter to recharge the battery.
The scale may continue to be used on AC power during charging.
Full charging time is approximately 10-12 hours. Battery life and recharge time will vary with use.
Over time, the operating time per each full charge will degrade. If the operating time is no longer
acceptable, the battery must be replaced.
When storing the scale for extended periods, the battery must be charged every 90 days to avoid
premature performance degradation.
CHARGING THE BATTERY