Step Parameter mark description Parameter mark
Operation manual
9
Normal display
Press
/A-ADJkey about 3 sec,into DZERO adjustment page
DZERO(Display Zero Adjust
1.Adjustment display zero with
or
key
2.Press
key enter data and into DSPAN adjustment page
DSPAN(Display Span Adjust
1.Adjustment display span with
or
key
2.Press
key enter data and return normal display
Step Parameter mark description Parameter mark
Operation manual
10
Normal display
Press
/A-ADJkey about 3 sec,into AZERO adjustment page
AZERO(Analog Output
Zero Adjustment page)
Value on EEPROM reset=0
1.Adjustment analog output zero with
or
or
key(±9999)
2.Press
key enter data and into ASPAN adjustment page
ASPAN(Analog Output
Span Adjustment page)
Value on EEPROM reset=0
1.Adjustment analog output span with
or
or
key(±9999)
2.Press
key enter data and return normal display
Appendix
Error Mark Analyze & Description
1 Input over error detect
Input signal over range(120%)
2 Input under error detect
Input signal under range(-20%)
3
Display over error detect
Display over range(19999)
4
Display under error detect
Display under range(-19999)
5 A/D Converter error
detect
1.
Input signal over range (180%)
2.
Inside circuit damage
Please moving input signal if still display ADER,please
6 EEPROM error detect
1.External interference when EEPROM read/write
2.EEPROM write over 100 million times(guarantee 10 years)
Please power reset,if still display E-
step:
1.
E-00 & No alternate display for inquire reset EEPROM
2.
Decide Yes with
or
key,press
display
3.EEPROM was reset,Please follow step 1~10 set again