4. Figure 7-46 will then appear. By
pressing F4 (OK), current adjustment
will start and the converter will output
a current corresponding to 0% of the
flow range. If it is fine, press F4(OK)
Figure 7-46
5. Figure 7-47 will then appear. When
executing the adjustment, select
“SET”, and then press F4 (ENTER).
Figure 7-47
6. A numerical value input display
(Figure 7-48) will then appear.
Measure the output current from the
converter and enter the current value
by the mA into the device. Once the
input has been completed press F4
(ENTER). The converter will start
adjusting to 4mA automatically, and
return back to the display as shown in
Figure 7-48. Verify that the current
output value is 4 mA, and terminate
the adjustment.
Figure 7-48
7. For 20 mA current output adjustment, follow the same procedures as shown in
above.