5. Advanced operation
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Enter the DICOM preferences submenu.
4. Select reset DICOM calibration to restore the original (not corrected) DICOM curve.
5.19.6 DICOM error threshold
About DICOM error threshold
The threshold to definetheDICOMcompliancecanbemodified in steps of 5% starting from 5 to 30%.
When the maximum deviation is not bigger than the selected threshold, the compliance check will be OK.
To set the DICOM error threshold:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Enter the DICOM preferences submenu.
4. Set DICOM error threshold as desired and confirm.
5.20 Imag e scaling
About image scaling
Enabling image scaling will multiply each individual pixel to one or more adjacent pixels so that the size
of the displayed image will be a multiple of the original image source video input signal.
Image scaling is only possible w hen t he resolution of your display’s video input signal
is less than or equal to half the maximum resolution of the display.
To enable/disable image scaling:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Image Source menu.
3. Enter the Scaling submenu.
4. Select Enabled/Disabled as desired and confirm.
5.21 Video input signals
About input signals
The available input signals for your display are:
DisplayPort 1 The input corresponding to the DisplayPort
connector.
DVI 1 The input corresponding to the DVI connector.
Automatic Selection The input is automatically selected.
To manually select a video input signal:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Image Source menu.
K5902035 MAMMO TOMOSYNTHESIS 07/12/2017