6. Important information
6. IMPORTANT INFORMATION
6.1 Safety information
General recommendations
Read the safety and operating instructions before operating the device.
Retain safety and operating instructions for future reference.
Adhere to all warnings on the device and in the operating instructions manual.
Follow all instructions for operation and use.
Electrical Shock or Fire Hazard
To prevent electric shock or re hazard, do n ot remove cover.
No serviceable parts inside. Refer servicing to qualied personnel.
Do not expose this apparatus to rain or moisture.
Type of protection (Electrical)
Equipment with external power supply: Class I equipment
Degree of safety (ammable anesthetic m ixture):
Equipment not suitable for use in the presence of a ammable anesthetic mixture with air or with oxygen
or nitrous oxide.
Non-patient care equipment
Equipment primarily for use in a health care facility that is intended for use where contact with a patient
is unlikely (no applied part).
The equipment may not be used with life support equipment.
The user is not supposed to touch SIP/SOPs and the patient at the same time.
Mission c ritical applications
We strongly recommend there is a replacement display immediate
ly available in mission critical applica-
tions.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equip-
ment (such as monitors). An activated electrosurgical generator may cause interference with them. The
interference can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection Equipment with external 24 VDC power supply
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC
(
) SELV power supply.
The medical approved DC (
) power supply must be powered by the AC mains voltage.
The power supply is specied as a part of the ME equipment or combination is specied as a ME
system.
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with
protective earth.
The equipment should be installed near an easily accessible outlet.
The equipment is intended for continuous operation.
(451920611391)K5903031 MDSC-2224 21/01/2014
47