F&E Equipment
Service manual V1.0 Gelovit 200
Autor : Fin/DW Stand 11.02.09 Seite 5 von 6
4.7 Readout of the hours of operation (device)
Switch on the device and wait until the device is ready to work. Press and hold the cycle
button. When the device has beeped 4 times, release the button. The device shows you
the operating hours now. The displayed value must be multiplied with 10.
4.8 Readout of the hours of operation (motor)
Switch on the device and wait until the device is ready.
Press and hold the timer button. When the device has beeped 4 times release the button.
Device shows the operating hours of the motor now. The displayed value must be multiplied
with 10.
4.9 Calibrating the sensors
Switch on the device and wait until the device is ready. Press and hold the T1&T2&T3 but-
tons. When the device has beeped 4 times, release the buttons.
With + / - buttons it is possible to set a correction value of ±0.9 ° C for the sensors. Upon
delivery, this value is 0. The device remains in the calibrating mode, please switch it off/on
the device to turn back to the user mode.
5. Known bugs
Currently no
6. Elimination of faults
Fault Cause Remedy
SOS sound display or cable to the display Call Customer
defective Service Dept.
E01 Gel sensor defective Sensor exchange,
E02 Lower temperature sensor defective PCB exchange
E03 Upper temperature sensor defective
E04 Rotary sensor defective Motor exchange,
PCB exchange
E05 Clock defective PCB exchange
E20 24V supply failure Check fuses,
Check power
supply at PCB
E30,31,32 engine brake failure PCB exchange